Over half of the breathalyzers tested by MHF Test Lab gave dangerously inaccurate readings, with some showing 0.0‰ even when the driver was over the legal limit. The test, done with Testfakta, highlights serious safety risks. Only fuel cell-based models like Dräger 4000 and Dignita AM-7000 showed reliable results, while cheaper semiconductor devices failed.